6-28
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Problem
Cause
Suggested Action
Detector Firmware Not
Supported
CPC firmware has not been
upgraded yet.
CPCs built before 2014 will need a firmware upgrade to
be compatible with the 3082 Classifier. The firmware
upgrades are available for download from the TSI
website. See
Detector Inlet
Flow Error
when connected to
Classifier
Impactor pressure drop is too
high.
When CPC in high-flow mode (1.5 L/min),
do not
use
0.0457 or 0.0508 impactors. If CPC is being used in
high flow mode, install either an inlet adapter or
0.071 impactor.
Cannot achieve desired
aerosol flow rate when flow
equalizer assembly is
installed
No impactor is installed.
An impactor must be installed on the inlet of the
classifier in order to achieve low sample flow rates when
using the flow equalizer assembly. See
Sheath Temp does not
match Cabinet Temp
Sheath flow is off.
This is expected. When sheath flow is off, the
temperature sensor in the sheath flow path (on the
flowmeter) experiences a 2 to 3°C heat-up due to circuit
electronics. An accurate sheath flow temperature
reading requires at least 2 L/min flow rate through the
sheath flow path.
3088 neutralizer is not on
but key switch is in the ON
position
3082 has been power cycled
with the key switch in the ON
position.
This is normal operation for regulatory purposes. Key
switch must be cycled from OFF position back to ON
position after a power cycle in order to enable
neutralizer.
3088 is not seated properly.
Top of 3088 should be flush with the top of the
classifier. Check for a gap. Pull out and reseat if
necessary.
Aerosol flow is negative
dP needs to be zeroed.
Zero the dP sensor. See
Inlet impactor is difficult to
insert
Lack of grease on internal 1-025
O-ring
Grease O-ring, see
3082 does not recognize a
USB flash drive
USB port has experienced over-
current.
Reboot the Electrostatic Classifier.
Cannot connect to 3082
from a PC using USB
PC is running Windows 8, and
Windows 8 driver not installed.
Install Windows 8 driver if available, or connect to PC
using Ethernet instead of USB. See
Connecting the Classifier to the Computer
3082 has not been power cycled
since the last time it was
connected to a PC. 3082 must
be power cycled before
connecting to a different PC.
Power 3082 off, then power back on again, then
connect USB to the PC.
Instrument is locked up because
PC has entered
sleep/standby/hibernate mode.
This is a known issue. It is recommended that the
standby mode on your PC be disabled if it is going to be
used with the classifier.
CPC was off at some point
during an SMPS scan.
This is a known issue. If CPC is powered off during a
scan, the classifier may lock up over time. To recover,
power cycle the classifier.
Classifier does not power
on automatically after soft
shut-down and AC power
reconnection
AC power reconnection occurred
less than 60 seconds after soft
shut-down.
If AC power is removed and reconnected less than
60 seconds after shutting down the instrument from the
front panel, the instrument will not recognize that AC
power has been reconnected. Wait 60 seconds before
reconnecting AC power or power on the classifier from
the front panel.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......