6-18
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Cleaning/Changing the DMA Dacron Screen
The Dacron screen (1030389) is located at the top of the annulus
assembly can become contaminated and contribute to arcing. If you
experience arcing problems, replace or clean the screen.
An extra Dacron screen was included with the DMA when shipped. Two
additional Dacron screens are included in the optional 3082 Maintenance
Kit 3082R-MAINT.
Cleaning/Replacing the Long DMA Screen
During cleaning procedures, refer to Figure 6-4 below.
Figure 6-4
Cleaning the
Long DMA Dacron Screen