5-34
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
The Logging Options screen allows you to set up data logging options.
Figure 5-19 below shows the Logging Options screen.
Figure 5-19
Electrostatic Classifier Logging Options Screen
To log data, follow these instructions:
1. On the Setup screen, touch
Logging Options
.
2. Touch
Enable Logging
to allow data to be logged (default setting).
Inverted data is logged to a .txt file. If
Enable Logging
is unchecked,
sample data will not be saved in any format.
3. To specify a
Delimiter
for the inverted logged data, touch
Tab
,
Comma
, or
Semicolon
. Tab delimiter is recommended to avoid
unintentional delimiting of any text in the Notes field which might use
commas or semicolons.
4. To specify a
Decimal Symbol
for the inverted logged data, touch
Dot
or
Comma
.
Note
: You cannot select comma for both the delimiter and the decimal
symbol. If you have specified a comma delimiter and then select
a comma decimal symbol, the delimiter changes.
5. Touch
Log
Raw Data
to enable the logging of raw data in addition to
the inverted data. A .TIM file containing raw counts will be saved in
addition to the .txt file (see next section for details about .TIM files).
6. Touch
Log to Internal Memory
to log the data to the internal memory,
or leave unchecked to log to external USB drive (preferred). An on-
screen message tells you how much logging memory is available on
the destination drive. If box is unchecked and an external drive is not
detected, the message will report a value of 0.0.
7. Touch
OK
to save your settings or
Cancel
to cancel them.
Note:
If
Log to Internal Memory
box is unchecked but no flash drive
is present when the start button is pressed, the classifier will
automatically check the box and switch to log to internal
memory.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......