Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
5. Using a Philips-
head screwdriver,
remove the four
screws holding the
flag assembly
housing in place
and save for reuse.
6. Lift the flag
assembly housing
out of the classifier
by placing one
finger on either
side.
7. Lift out the flag assembly housing and rest it
on the top of the instrument.
Note
: The housing is tethered and cannot be
removed completely.