Maintenance, Service, and Troubleshooting
6-5
5. Remove the thicker, 1-207 face-seal O-ring (6006680) from the bottom
face of the impaction plate.
6. Remove replacement O-rings and vacuum grease from the 3082
Accessory Kit. Follow the instructions in the Greasing O-rings section
below and smear a small amount of vacuum grease onto the
replacement O-rings.
C a u t i o n
The 1-207 O-ring (6006680) is made of conductive material to ensure
that the impactor is grounded to the classifier chassis. It should
not
be
replaced with a substitute 1-207 O-ring.
If you set the O-rings on a dirty surface, they will pick up debris. Perform
a visual inspection and remove debris with a clean wipe.
7. Slot new O-rings into the correct grooves in the impactor body.
To replace the 1-025 O-ring (6006389) from the classifier inlet, follow
these instructions:
1. If an impactor or inlet adapter is installed in the inlet of the classifier,
remove it.
2. Locate the O-ring.
Note
: It may be helpful to use a light.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......