SYMPTOM
PROBABLE CAUSE
REMEDY
5.
Unit dispenses only partially, or
not at all.
The manifold dispense needles are
clogged, or have cracked glass.
Clean or replace needles. See
Sections 4.6-4.7.
The Wash reservoir inlet filter is
clogged.
Backflush or replace filter. See
Section 4.3.
Tubing is kinked or obstructed or is
not connected properly.
Check connections, replace worn
out tubing. See Section 1.6.
6.
Instrument overfills wells.
Clogged dispense or aspirate
needles.
Clean aspirate and dispense
needles. See Sections 4.6-4.7.
The aspiration pump or Waste
reservoir tubing are kinked,
obstructed. or connected
improperly.
Check connections, replace worn
out tubing, See Section 1.6.
The cap on the Waste reservoir is
loose.
Tighten cap.
Manifold needles are not parallel to
the well strips.
Adjust the manifold so that it is
level. See Section 1.5.1.
7.
Waste reservoir sensor does
not register “FUL” when
submerged in liquid.
Accumulated residue preventing
float switch from functioning.
Clean float switch (on probe
attached to inside of waste bottle
cap.
8.
Manifold hitting plate after
washing several rows properly.
Tubing length incorrect causing
pulling/pushing of manifold out of
alignment.
Reposition tubing to the proper
lengths. See Section 4.8.
9.
Manifold aspirate (metal)
needles are bent.
Dropping of the manifold or
repackaging with proper padding.
Carefully push the needle/needles
back into straight alignment.
10. Waste bottle collapses.
The aspirate needles in the
manifold are clogged.
Clean the manifold needles with
supplied tool.
5.2
TROUBLESHOOTING TABLE (CONTINUED)
Multiwash III
32
Содержание 8441 series
Страница 1: ...8618 01 E MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL ...
Страница 5: ...Multiwash III THIS PAGE INTENTIONALLY LEFT BLANK vi ...
Страница 14: ... FINAL ASPIRATION WELL POSITION TOP OF WELL ASPIRATION POSITION Multiwash III 9 ...
Страница 40: ...5 4 USER TROUBLE SHOOTING NOTES Multiwash III 35 ...