
4.7
REPLACING STANDARD MANIFOLD DISPENSE NEEDLES
Note: A needle service tool—included with the optional Maintenance Kit and in the
Replacement Needle Kit—is required to replace manifold dispense needles.
Insert the
needle service tool under the dispense needle collar. Pry the dispense needle upward with the
tool. This will loosen the needle from its press fit with the manifold. Place the new needle into
the open end of the needle service tool and firmly insert it into the manifold.
Care should be exercised to prevent breakage of the precision glass insert.
Multiwash III
26
Содержание 8441 series
Страница 1: ...8618 01 E MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL ...
Страница 5: ...Multiwash III THIS PAGE INTENTIONALLY LEFT BLANK vi ...
Страница 14: ... FINAL ASPIRATION WELL POSITION TOP OF WELL ASPIRATION POSITION Multiwash III 9 ...
Страница 40: ...5 4 USER TROUBLE SHOOTING NOTES Multiwash III 35 ...