4.8
INSTALLING MANIFOLD TUBING
Use these step-by-step instructions to install the tubing for proper functioning of the manifold,
remembering that the dispense and aspirate needles on the manifold are fragile.
Multiwash III
3
IDENTIFY EXISTING OR NEW TUBING, WASTE/MANIFOLD ASSEMBLY "A" AND TUBING SYRINGE/MANIFOLD
ASSEMBLY "B" AS SHOWN ABOVE.
THREAD THE CLEAR FITTING ON TUBING ASSEMBLY "A", THRU THE CONDUIT FROM THE REAR OF THE
UNIT AS SHOWN.
THREAD THE CLEAR FITTING ON TUBING ASSEMBLY "B", THRU THE CONDUIT FROM THE FRONT OF THE
UNIT
AS SHOWN.
2
CONDUIT
CLEAR FITTING
THREAD THRU
FROM FRONT
WASTE/MANIFOLD
TUBING ASSEMBLY “A”
CLEAR FITTING
THREAD THRU
FROM BACK
2
SYRINGE/MANIFOLD
TUBING ASSEMBLY “B”
WHITE FITTING (REF.)
3
(continued on next page)