4.6
CLEANING MANIFOLD NEEDLES
A needle cleaning wire is supplied with each manifold, and may be used to unclog aspirate or
dispense needles which have been plugged by particulates or dried salt deposits. The
standard manifold dispense needles are a composite assembly with an outside sheath of
plastic, housing an inner precision bore glass capillary tube. Avoid bending the assembly when
handling. The cleaning wire can be used when rinsing fails to clear a bore.
Multiwash III