5
TROUBLESHOOTING
5.1
RE-INITIALIZING THE STORED PROGRAMS
To re-initialize the Multiwash III's program settings, hold the Down Arrow key on the keypad as
you turn the power switch on. This will return all user defined programs back to the factory
settings!
5.2
TROUBLESHOOTING TABLE
The following checklist may help resolve problems in the operation of this equipment. Any
problems that cannot be resolved by following the procedures below should be referred to
qualified service personnel.
Multiwash III
SYMPTOM
PROBABLE CAUSE
REMEDY
1.
Instrument is “dead”. Display is
blank, the beeper does not
sound when keys are pressed
and motors do not operate.
AC power not getting to the
instrument. Check Power Supply
cable.
Blown fuse.
Connect to Power Supply Cable.
Instrument setup for the wrong
voltage.
Replace fuse with proper type. See
Section 4.9.
Change voltage selector to the
correct position. See Section 1.3.
2.
Instrument doesn’t respond to
key presses.
Invalid data in instrument’s stored
programs. Caused by voltage
spikes or fluctuations.
Hold “Down Arrow” key while
turning on power to restore
operation. A line conditioner or
surge protector needs to be used.
See Section 1.2.
3.
Manifold hits plate when
aspirating rows.
The manifold is not calibrated to the
proper position within the wells.
Calibrate horizontal manifold
position. See Section 1.5.2.
4.
Instrument aspirates wells only
partially, or not at all.
The manifold aspirate needles are
partially or completely clogged.
Clean needles with supplied tool.
See Section 4.6.
Tubing is kinked or obstructed or is
not connected properly.
Check connections, replace worn
out tubing.
The waste reservoir cap is loose, or
fittings on cap are loose.
Tighten Waste reservoir cap, on
bottle and all fittings.
Manifold aspiration needles are not
optimized for final aspiration.
Adjust Horizontal manifold position
Section 1.5.2. Adjust vertical
manifold position Section 1.5.3.
Verify correct plate selection
Section 3.3.8.
31
Содержание 8441 series
Страница 1: ...8618 01 E MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL ...
Страница 5: ...Multiwash III THIS PAGE INTENTIONALLY LEFT BLANK vi ...
Страница 14: ... FINAL ASPIRATION WELL POSITION TOP OF WELL ASPIRATION POSITION Multiwash III 9 ...
Страница 40: ...5 4 USER TROUBLE SHOOTING NOTES Multiwash III 35 ...