1-72
(a) Z-path method
This type of method is effective for application of measurement on the larger pipe or
measurement for attenuating fluid against ultrasonic. Because of transducers distance is
relatively shorter, so please select this method when V-path method installation is
impossible.
Fig. 1.2.8-19 Z-path method
(b) V-path method
This method is typical. Please select this method as primary.
Fig. 1.2.8-20 V-path method
(c) W-path method
This method may be effective for smaller pipe measurement. Please select this method
when you have any problems by V or Z-path method.
Fig. 1.2.8-21 W-path method
Pipe
Transducer
Transducer
Pipe
Transducer
Transducer
Pipe
Transducer
Transducer