3-30
• Flow values do not vary.
Q; Does “R1/R2” marks appear?
A; There is no echo receiving.
The receiving signal of enough strength is not obtained.
See Chapter “
3.6.2 Measurement
(p.3-31)”.
Q; Does “D1/D2” marks appear?
A; The obstacle has been detected.
The receiving signal is unstable. Or, the appearance of the flow has changed rapidly.
If frequency of occurrence is high, see Chapter “
3.6.2 Measurement
(p.3-31)”.
Q; Does “E1/E2” marks appear?
A; The setting is improper. Therefore, a correct measurement preparation cannot be
done.
Confirm the setting. Especially, confirm the coverage of the sensor.
See Chapter “
2.2.4 Advanced setting operation (3) Sensor condition
(p.2-19)”.
For example : DN 500 mm, UP50AST, V-path method --- improper
DN 1000 mm, UP10AST, V-path method --- improper
DN 5000 mm, UP04AST, V-path method --- improper
• USB memory is not recognized.
A part of USB memory cannot be recognized. Pull out and put into USB memory. Check
other makers’ USB memory.
It doesn’t correspond to the card reader.
• Battery is not charged.
The temperature range of the battery that can be charged is Approx.0
~
50°C. The charge
stops automatically outside this range. See Chapter “
1.2.3 Procedure for start up (4)
Battery Charge
(p.1-17)”.
Содержание UFP-20
Страница 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Страница 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Страница 8: ...1 Installation ...
Страница 9: ......
Страница 17: ......
Страница 18: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 130: ...2 Operation ...
Страница 131: ......
Страница 182: ...3 Other ...
Страница 183: ......
Страница 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...
Страница 195: ...3 10 1 Adaptor A 2 Adaptor B 3 Knob Fig 3 2 4 5 Fixture 1 Fig 3 2 4 6 Fixture 2 Fig 3 2 4 7 Z path method adaptor ...
Страница 196: ...3 11 4 Large Transducer Kit Fig 3 2 4 9 Fixture 1 Fig 3 2 4 10 Fixture 2 Fig 3 2 4 8 Large transducer ...