1-112
4C-d. Rewrap
Rewrap the paper around the pipe at the intended flow measure point.
Fig. 1.2.9-61 Fix gauge paper
4C-e. Extend the line
Using a pencil or marking pen, mark the pipe at both edges of the paper at the overlap and at
the crease by using a pencil or marking pen to extend the lines outward from the paper onto
the pipe.
Fig. 1.2.9-62 Mark the extend lines
(1)
(2)
Adhesive tape
Crease
Extend Marking Line
Extend Marking Line
Required F-Dist
Adjust both edges
Fix gauge paper
using adhesive tape.