1-89
Fig. 1.2.9-18 Flow of medium transducer installation
Note: The distance between transducers needs to be determined prior to
transducer installation. Please refer to section 1.2.6 ~ 1.2.8 to
determine this distance before proceeding with the steps described
below for installation.
3-a.
Slide distance setting
3-b.
Set mounting fixture onto the pipe
3-c.
Attach Z-path method adaptor to mounting fixture
3-d.
Add couplant and set transducers to mounting fixture
3-e.
Set cables with the transducers and the main unit
3-f.
Let’s start measurement