
2-24
(6) Volume correction
Change zero shift, span correction and so on.
Fig. 2.2.4-11 Volume correction menu
1. Zero shift
Change zero shift value. Unit is defined by flow/mass/energy unit.
Table 2.2.4-15 Range of zero shift
Default
min.
max.
0.000
-99999.0
999999.0
2. Span correction
Change span correction value.
Table 2.2.4-16 Range of span correction
Default
min.
max.
1.000
0.000
20.000
The correction value is calculated below equation,
q
=
a
x
v
+
b
where,
“
q
” is correction value, “
a
” is span correction, “
v
” is without correction value and “
b
” is
zero shift.
V o l u m e C o r r e c t i o n
1 :
Z e r o S h i f t
0 . 0 0 0 m
3
/ s
2 : S p a n C o r r e c t i o n
1 . 0 0 0
3 : Z e r o C u t
0 . 0 0 0 m
3
/ s
4 : V o l u m e C o r r e c t i o n
R e y n o l d s
# 0 1 : 1 0 0 / S t e e l / 1 M
B a c k
S e l e c t
Содержание UFP-20
Страница 1: ...Doc No KF08 002L Portable Ultrasonic Flowmeter UFP 20 Installation Operation Manual ...
Страница 3: ... 2 The attachment of warning label is as follows Fig 1 2 Rear view of main unit Warning label ...
Страница 8: ...1 Installation ...
Страница 9: ......
Страница 17: ......
Страница 18: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 130: ...2 Operation ...
Страница 131: ......
Страница 182: ...3 Other ...
Страница 183: ......
Страница 193: ...3 8 3 2 4 Dimensions 1 Main Unit Fig 3 2 4 1 Main Unit Hand Strap ...
Страница 195: ...3 10 1 Adaptor A 2 Adaptor B 3 Knob Fig 3 2 4 5 Fixture 1 Fig 3 2 4 6 Fixture 2 Fig 3 2 4 7 Z path method adaptor ...
Страница 196: ...3 11 4 Large Transducer Kit Fig 3 2 4 9 Fixture 1 Fig 3 2 4 10 Fixture 2 Fig 3 2 4 8 Large transducer ...