3-a. Transducer distance setting
Set slide distance as indicated value which is calculated by main unit +50mm.
Please note that proper transducer distance is calculated by manually for Z-path method
installation.
Fig. 1.2.9-19 Message of mounting information Fig. 1.2.9-20 Set slide distance on fixture
M e s s a g e
M o u n t i n g I n f o r m a t i o n
CH1
Distance XX.X mm
Sound-path Z-path method
Transducer UP10AST
Note1:
Slide distance = indicated di 50 mm
Note2:
Transducer distance indicated on Fig.1.2.9-19 is different from set distance as
Fig.1.2.9-20 in case of using Z-path method adaptor.
Slide distance
Slide
Slide