4B-d. Rewrap
Rewrap the paper around the pipe at the intended flow measure point.
Fig. 1.2.9-46 Fix gauge paper
4B-e. Extending the line
Using a pencil or marking pen, extend the crease lines outward from each edge of the paper
onto the pipe (at points (1), (2) in Fig. 1.2.9-47). With the outside edge of F-Dist which is to be
positioned at one edge of the gauge paper, measure and mark off the F-Dist to the other
transducer holder as shown in Fig. 1.2.9-47.
After marking, remove the gauge paper.
Fig. 1.2.9-47 Mark the extend lines
Adhesive tape
(1)
(2)
Crease
Extend Marking Line
Required F-Dist
Extend Marking Line
Adjust both edges
Fix gauge paper
using adhesive tape.