Service Procedures
Doc. 031880-04 1/12
71
2. Go to the
DETAIL
screen (see the examples in Figure 67 and
Figure 68) and set the following pressure limits:
Low limit = 1.38 MPa (200 psi)
High limit = 20.6 MPa (3000 psi)
NOTE If you are installing a Dionex RFC-30, the pressure
limits listed above must be set. These limits are
recommended for a Dionex RFC-10, but not
required.
Figure 67. Example DX-320 Detail Screen
Figure 68. Example DX-500/DX-600 Pump Detail Screen
NOTE When pump operation is controlled by Thermo
Scientific Dionex chromatography software,
pressure limits can also be set from the software.
3. If you are installing a Dionex RFC-10, go on to Section B.6. If you
are installing a Dionex RFC-30, go on to Section B.7.
100
OUTPUT 0.05uS
LOCAL
OVEN TEMP
0 PSI
5.00 mL/MIN
TTL1
TTL2
RLY1
RLY2
0
0
0
0
RANGE
P-POINT L= 0 R=0
LIMIT
DIRECT CNTRL
OFFSET 189.13 uS
C READY
uS
SRS OFF mA
TEMP COMP 1.7
200 - 3000 PSI
TOTAL 189.18 uS
25
LOAD
0.30
59
LOCKED RMT
COLUMN
DETAIL SCREEN
100.0 %A
RLY1
RLY2
TTL1
TTL2
0
0
0
0
LOAD
200-3000 PSI
0.0 %B
0.0 %C
0.0 %D
LIMIT
DIRECT CNTRL
PSI
mL/MIN
SAMPLE
A