Dionex Reagent-Free Controller
58
Doc. 031880-04 1/12
4. Reconnect the following EGC holder liquid lines, if they were
disconnected:
•
Connect
TO PUMP/DAMPER
to the pump outlet.
•
Connect the
WASTE, GAS SEPARATOR
to the gas separator waste
tube assembly installed on the waste container.
5. Turn on the pump and set the flow to the rate recommended for your
application. Pump deionized water through the CR-TC, for at least 10
minutes.
6. Turn off the pump and disconnect the coupler.
7. After hydrating the CR-TC,
•
Reconnect
TO INJ VALVE IN - P
to port
2
on the injection valve.
•
Reconnect
TO SRS/AES REGEN OUT
to the
REGEN OUT
port on
the suppressor.
8. Connect the cable from the CR-TC to the
CR-TC
cable from the
Dionex RFC and connect the cable from the EGC to the Dionex RFC
EG-1
cable. Push two cables together and twist the locking ring to
connect them.
9. Turn on the power to the Dionex RFC.