Dionex Reagent-Free Controller
28
Doc. 031880-04 1/12
Setting Up the Dionex RFC-30 for Manual Current Entry
If you do not want the Dionex RFC-30 to automatically calculate and set
the suppressor current as described in the previous section, you can turn
off automatic calculation. To do this, select one of the following
suppressor types, depending on the type installed:
SRSXCALC
(for a
Dionex AES suppressor—no calculation) or
SRSXCALC
(for an SRS
suppressor—no calculation). See the example below.
1. Set the EGC type, flow rate, and eluent concentration as described in
the previous example for automatic calculation.
2. Then, go to the suppressor type screen.
Figure 27. Suppressor Type Screen
3. Press
ENTER
. The
SUPP TYPE =
entry field is highlighted.
Figure 28. Selecting the Suppressor Type
4. Press
and select either
SRSXCALC
or
AESXCALC
.
Figure 29. AESXCALC Type Selected
5. Press
ENTER
.
6. Press
three times to go to the suppressor current screen.
Figure 30. Suppressor Current Screen
7. Press
ENTER
.
8. Press
to select the desired suppressor current and then press
ENTER
.
Current limits
for the AES
suppressor