Dionex Reagent-Free Controller
52
Doc. 031880-04 1/12
11. Connect
EGC IN
to
the EGC
INLET
Figure 47. EGC Inlet Connection
12. Locate the yellow
0.5 mL/min, 1000
psi (7 MPa)
pressure restrictor
coil (P/N 053765)
in the Dionex
RFC-30 Ship Kit
(P/N 060299).
Connect this coil
to the EGC
OUTLET
Figure 48. EGC Outlet Connection for
Conditioning the EGC
NOTE This is a temporary connection, which is used only
during the EGC conditioning procedure described in the
next section.
13. Turn over the EGC (fittings facing down).
Shake the EGC
vigorously, and tap it with the palm of your hand 10 to 15 times.
Check that all trapped bubbles are dislodged.
INLET Port
EGC IN
Pressure Restrictor
Coil
OUTLET
Port