Dionex Reagent-Free Controller
56
Doc. 031880-04 1/12
6. Lift the holder out of the
rack, turn it upside
down, and set it on the
workbench.
7. Grasp the top of the
CR-TC and lift up to
disconnect it from the
holder.
8. Disconnect the liquid
lines from the four ports
on the CR-TC (see
Figure 53).
Figure 53. Disconnecting CR-TC Liquid Lines
9. Remove the CR-TC from the holder.
10. Remove the plugs from the ports on the new CR-TC. Reconnect the
liquid lines that were disconnected in Step 8 to the corresponding
ports on the new CR-TC.
11. Check that none of the
lines are caught under
the CR-TC and then
push the CR-TC onto
the metal stud inside the
EGC holder. Push down
until the CR-TC snaps
in place (see Figure 54).
Figure 54. CR-TC Installed in Holder
Disconnect liquid
lines from 4 ports