65
Table 3-4
Measurement Function
Setting
Trigger Delay Time
DCV/DCI
PLC >= 1
1.5 ms
PLC < 1
1.0 ms
Ω2 and Ω4
(PLC >= 1)
100Ω
to
100kΩ
1.5 ms
1 MΩ
15 ms
10 MΩ
to
100 MΩ
100 ms
Ω2 and Ω4
(PLC < 1)
100Ω
to
100kΩ
1.0 ms
1 MΩ
10 ms
10 MΩ
to
100 MΩ
100 ms
ACV/ACI
(Remote Interface/
External Trigger/
Single Trigger
3 Hz
7.0 s
20 Hz
1.0 s
200 Hz
600 ms
ACV/ACI
(Front Panel /
Auto Trigger On)
3 Hz
1.5 s
20 Hz
200 ms
200 Hz
100 ms
Frequency/Period
Remote Interface /
External
1.0 s
Front Panel /
Auto Trigger ON
0 s
Capacitance
1 nF to 10 mF
0 s
How to specify a delay time
User can set the delay time from either the front panel operations or the remote
interface operations.
Содержание DL-1060
Страница 1: ...INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 1060 DL 1060G DL 1060VG DL 1060R DL 1060VR B71 0180 31...
Страница 91: ...85 3 4 10 Self test Notice Do not use SELF TEST since it is used only for an after sales service operation...
Страница 146: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...