3.1.4 Threshold Resistance (Continuity)
Definition
When testing continuity, the beeper goes off when the measured resistance is less
than the threshold resistance. The threshold resistance can be set to any value
between 1
Ω and 1000Ω.
Default
The factory default for continuity threshold resistance is 10Ω.
Your selection is stored
in a volatile memory and the default value will be restored after the meter has been
turned off.
How to set the threshold resistance
You can change the threshold resistance only through the front panel as shown in
Figure 3-8
.
1.
Press CONT button and then CONFIG button.
2.
Press ENTER button to access the threshold resistance settings.
3.
Use
◁
and
▷
buttons to move through the digits and lower or increase the
number to your desired value by using
△
and
▽
buttons.
4.
Press ENTER button again to confirm the value.
Figure 3-8