102
Appendix
A. Specification List
DC Characteristics
Accuracy ± (% of r % of range) [1]
Function
Range
[
2
]
Resolution Input resistance 1 Year (23
℃
± 5
℃
)
DCV
(DC Voltage)
100.0000 mV
0.1 μV
10
MΩ
0.008 + 0.0045
1.000000 V
1 μV
10
MΩ
0.009 + 0.001
10.00000 V
10 μV
10
MΩ
0.012 + 0.002
100.0000 V
100 μV
10
MΩ
0.012 + 0.002
1000.000 V
1 mV
10
MΩ
0.02 + 0.003
Function
Range
[
2
]
Resolution
Shunt
Resistance
1 Year (23
℃
± 5
℃
)
DC Current
10.00000 mA
10 nA
5.1
Ω
0.05 + 0.02
100.0000 mA
100 nA
5.1
Ω
0.05 + 0.01
1.000000 A
1
μ
A
0.1
Ω
0.15 + 0.02
3.00000 A
10
μ
A
0.1
Ω
0.2 + 0.03
10.00000 A
10
μ
A
0.005
Ω
0.25 + 0.05
Function
Range
Resolution
Test Current
1 Year (23
℃
± 5
℃
)
Resistance
[3]
100.0000 Ω
100
μΩ
1 mA
0.02 + 0.005
1.000000 kΩ
1 mΩ
1 mA
0.02 + 0.002
10.00000 kΩ
10 mΩ
100 μ
A
0.02 + 0.002
100.0000 kΩ
100 mΩ
10 μ
A
0.02 + 0.002
1.000000 MΩ
1 Ω
1
μ
A
0.02 + 0.004
10.00000 MΩ
10 Ω
0.1
μA
0.1 + 0.004
100.0000 MΩ
100 Ω
0.1
μA
(//
10 MΩ
)
1.5 + 0.005
Diode Test
1.0000 V
10 μV
1 mA
0.02 + 0.02
Continuity
1000.00
Ω
10 mΩ
1 mA
0.02 + 0.03
[1] Specifications are for 2-hour warm-up condition, 10PLC and they
’
re relative to calibration
standards.
[2] 20
%
over range on all ranges except 1000Vdc and 3A/10A range.
[3] Specifications are for 4-wire or 2-wire when a Null operation is used.
Содержание DL-1060
Страница 1: ...INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 1060 DL 1060G DL 1060VG DL 1060R DL 1060VR B71 0180 31...
Страница 91: ...85 3 4 10 Self test Notice Do not use SELF TEST since it is used only for an after sales service operation...
Страница 146: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...