6
1.2.2 To connect the power
Check the power-line voltage on the rear panel to see if voltage setting is correct for
your area. Change the voltage correctly or replace a new fuse if the voltage is not
correct or the fuse is broken. Please follow the steps below.
1.2.2.1 To convert the voltage
In some areas, the power supplied voltage is 240Vac or 120Vac; in
others, the power supplied voltage is 220 Vac or 100 Vac. Please
refer to your local power supplied voltage to see if you have the
right setting.
Before changing the setting, make sure the multimeter is
disconnected from the AC power. An incorrect voltage setting may
cause severe damage to your instrument. The power cord supplied
with DL-1060 contains a separate ground wire for use with
grounded outlets. When proper connections are made, instrument
chassis is connected to power line ground through the ground wire
in the power cord. Failure to use a grounded outlet may result in
personal injury or death due to electric shock.
Suppose your multimeter is under the AC 220V condition and you want to convert it to
100V. Follow the following steps to change the voltage setting.
【
Step 1
】
Verify that the DMM is disconnected as shown in
Figure 1-9
.
Figure 1-9
WARNING
!