32
2.8.2.1 2-Wire RTD Measurements
How to measure temperature with 2-Wire RTD
The following
Figure 2-10
shows theory diagram of 2-Wire RTD measurement.
1.
Insert a specified adapter into the front terminals.
2.
Press TEMP button.
3.
Configure sensor type and unit by using the CONFIG button (refer to 3.1.8
) or skip this step to use the default settings.
4.
Observe and take readings on the display.
Figure 2-10
2.8.2.2 3-Wire RTD Measurements
How to measure temperature with 3-Wire RTD
The following
Figure 2-11
shows theory diagram of 3-Wire RTD measurement.
1.
Insert a specified adapter into the front terminals, and connect the low thermal
patch leads to the adapter.
2.
Press TEMP button.
3.
Configure sensor type as you would be with 4-wire RTD and unit by using the
CONFIG button (refer to 3.1.8
) or skip this step to use the
default settings.
4.
Observe and take readings on the display.
※
Note: When users do the 3-wire RTD measurement the input LO and sense LO
have to be connected in short condition.
※
Note: “Source” means RTD Probe or RTD Simulator.
Содержание DL-1060
Страница 1: ...INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 1060 DL 1060G DL 1060VG DL 1060R DL 1060VR B71 0180 31...
Страница 91: ...85 3 4 10 Self test Notice Do not use SELF TEST since it is used only for an after sales service operation...
Страница 146: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...