126
For Medical Staff
Table a) – Manufacturer’s declaration – electromagnetic emissions – for IEC 60601-1-2:2014 (EN 60601-1-2:2015)
Manufacturer’s declaration - electromagnetic emissions
The TERUFUSION Syringe Pump Type SS3 is intended for use in the electromagnetic environment specified below. The
customer or the user of the TERUFUSION Syringe Pump Type SS3 should assure that it is used in such an environment.
Emissions test
Emissions class and group
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class [A]*
Harmonic emissions
IEC 61000-3-2
Class [A]
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies
* The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If
it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radiofrequency communication services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
Table b) – Manufacturer’s declaration – electromagnetic immunity – for IEC 60601-1-2:2014 (EN 60601-1-2:2015)
Manufacturer’s declaration - electromagnetic immunity
The TERUFUSION Syringe Pump Type SS3 is intended for use in the electromagnetic environment specified below. The
customer or the user of the TERUFUSION Syringe Pump Type SS3 should assure that it is used in such an environment.
Immunity test
Immunity test level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to 2.7 GHz
See Table c) for immunity test level to RF Wireless
communication equipment
Electrical fast transient / burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input / output lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Conducted RF
IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz outside ISM bands
6 Vrms 150 kHz to 80 MHz in ISM bands
Power frequency (50-60 Hz) magnetic field
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions and voltage variations on
power supply input lines
IEC 61000-4-11
0% U
T
for 0.5 cycle
0% U
T
for 1 cycle
70% U
T
for 25/30 cycles
0% U
T
for 250/300 cycles