Troubleshooting
When installing a syringe: Slider displacement
Occurrence
Cause
Action
• LCD: [Place the syringe.]
Slider displacement icon
flashes.
• The slider has been displaced.
• Place the syringe correctly.
(See page 26.)
Maintenance timer
Occurrence
Cause
Action
• The Maintenance timer
indicator
lights up.
• It is time to perform
maintenance.
• Contact TERUMO trained service technicians.
AC power detector function
Occurrence
Cause
Action
• The pump cannot be turned on
when using AC power.
• The AC icon
is not
displayed even after the AC
power cable is connected.
• Check the connection of the AC power cable to the
pump and the power supply. (See page 23.)
• The AC power cable is
damaged.
• The AC power cable is
internally disconnected.
• Replace the AC power cable.
Battery Replacement Time/Failure alarm
Occurrence
Cause
Action
• The battery operation time is
short even after fully charged.
• The battery icon
is displayed.
• The battery icon does not
display 5 bars after charging
for 8 hours or more.
• The battery has deteriorated
or is damaged.
• Replace the internal battery.
•
and the battery charge
level appear alternately after
more than 8 hours of charging.
• The sub-battery has
deteriorated or is damaged.
• Replace the sub-battery.