17
Precautions
Cautions
• This product should only be used with a sufficient power supply. [If the power supply is not sufficient, the internal battery
will be used for the operation, which may result in no power supply in an emergency situation.]
• When using devices (mobile phones, radio devices, radio knives, defibrillators, etc.) that emit an electromagnetic wave
within the area of this product, they should be used as far away as possible and you should verify the normal operation of
this product in the configuration it is used. In addition, this product should use a power supply of a system isolated from
these devices and should obtain secure earth connection. [Malfunction in this product may result in critical harm to the
patient.]
• When used in the vicinity of a radio knife (Medical radio knives are surgical
equipment for incision and coagulation by high energy radiofrequency
current), check the following before use.
(1) Radio knives have a different level of high-frequency noise emission
depending on the types, and using old models (vacuum tube gap type)
in combination should be particularly avoided as the noise levels from
those are higher.
(2) The distance from the radio knife cord (knife holder, knife cord and
return electrode cord) and radio knife body to this product should be kept
to a minimum of 25 cm.
(3) The radio knife and this product should be operated by a power supply
from a separate system, and both should be securely earthed.
• When connecting this product with other medical equipment or network system, check its conformance to IEC 60601-
1:2005+A1:2012 (EN 60601-1:2006+A1:2013) and IEC 60601-1-2:2014 (EN 60601-1-2:2015) prior to use in order to ensure
the system safety.
• When connecting this product with a network system or other equipment, for safety, checking the manufacturer’s
specifications of the equipment is recommended. In addition, use EMI compliant products for the connection cable.
• When using the external communication function, pay particular attention, as it is more susceptible to the effect from a
radio knife, mobile phone, radio device, defibrillator, etc. Regularly check that this product is operating normally.
• If connecting with a nurse call, always check in advance with the electrical contractor for compatibility, and connect with a
system for which the connection function has been confirmed. [The functions regarding connection are not guaranteed.]
(TE-SS732 only)
• If connecting with a buzzer or lamp instead of a nurse call, it should be used with a capacity of 12V DC, 1A or less.
(TE-SS732 only)
• When using the wireless LAN communication function, the effect on other equipment by radio interference should be
considered. [It may affect other equipment.] (TE-SS830 only)
• When connecting this product with a network, the settings of this product and the network system need to be compatible.
Contact a vendor specialist for the correct settings and allow only the system administrator to perform settings. [Failure
to connect with correct settings may result in interfering with the original functionality or performance of this product and
affecting the network system.] (TE-SS830 only)
• Connection to a network including other devices may cause unpredicted and unacceptable risks for the patient, the user,
or a third person. Be sure to identify, analyze, evaluate, and control those risks.
• Subsequent changes to the IT-network could introduce new risks and require additional analysis.
- Changes to the IT-network include:
(1) Changes in the IT-network configuration.
(2) Connection of additional items to the IT-network.
(3) Disconnecting items from the IT-network.
(4) Update of equipment connected to the IT-network.
(5) Upgrade of equipment connected to the IT-network.
• When using the IV pole, follow the instructions for the IV pole.
Medical
electrical
equipment
Power supply for
medical electrical
equipment
Keep 25 cm or more
Power supply from a different system
Power supply
for radio knife
Knife holder
Return electrode
Radio knife