124
Device Characteristics
Occlusion Characteristics
Typical occlusion detection ability data are occlusion pressure, Occlusion alarm occurrence time, and post-occlusion removal
bolus volume. The data below is a representative example measured for flow rates of 1 mL/h and 5 mL/h, with a three-stage
occlusion pressure setting value (1, 6, 10), and a new TERUMO syringe (50 mL). The measured values may vary depending on the
test conditions.
* When the 1m extension tube (internal volume: 1.0 mL) is used.
Note
• The Occlusion alarm occurrence time with solution delivery at the minimum flow rate (0.01mL/h) will be approx. 100 times of
those at the flow rate 1mL/h.
Flow rate
(mL/h)
Occlusion
pressure
setting value
Occlusion pressure
Time to activate an
Occlusion alarm time
(min)
Bolus volume
(mL)
(kPa)
(kgf/cm
2
)
1
1
10
0.10 68.2 0.04
6
62
0.64 92.3 0.54
10
113
1.16 140.6 0.63
5
1
10
0.10 2.0 0.06
6
62
0.63 18.3 0.46
10
112
1.14 25.2 0.69
Содержание TERUFUSION TE-SS730