120
Specifications
Available
Functions
Drug/patient
information display
function
Displays drug information such as dilution, patient information including
patient name, age, gender, weight, height, etc. The dilution and weight can be
changed.
Dilution setting
function by amount and
solution
Selects how to specify the dilution from either Dilution or Amount/Solution.
Interval/Intermittent
dose function
(TE-SS830 only)
The specified dosage rate is delivered at the specified regular intervals.
Multi-step dose
function
(TE-SS830 only)
The speed of solution delivery increases or decreases at the specified levels
(up to 10 levels).
Delayed Start dose
function
(TE-SS830 only)
Solution delivery starts after the specified time has elapsed.
Link function (Switch
function A, B)
(TE-SS830 only)
Switch function A: Operations of two pumps are linked. When a Completion
alarm occurs on the first pump, the second pump starts
solution delivery.
Switch function B: Interlocks two pumps. Starts solution delivery of the first
pump, and then starts solution delivery of the second pump
at the predetermined time. After solution delivery of the
second pump is started, each pump changes the solution
delivery speed according to respective settings. The
interlock of the pumps finishes upon completion of the set
solution delivery or at alarm timing.
Come and See ME
function
(TE-SS830 only)
Automatic call function after a specified time elapses.
Volume delivered unit
select function
When the dose rate is set in mass units, the volume delivered unit can be
changed to mass units.
Micro notation select
function
Switches “
μ
g” to “mcg” to be displayed in the drug library mode,
μ
g/kg/min
mode or mg/kg/h mode.
Use conditions
Temperature: 5 to 40°C
Humidity: 20 to 90% RH (non-condensing)
Atmospheric pressure: 70 to 106kPa (700 to 1060hPa)
Storage conditions*
Temperature: -20 to 45°C
Humidity: 10 to 95% RH (non-condensing)
Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
Transport conditions
Temperature: -20 to 60°C
Humidity: 10 to 95% RH (non-condensing)
Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
Power supply
AC100 to 240V, 50 to 60Hz
Battery: Lithium ion battery
• Continuous use period: Approx. 12 hours (applies when solution is delivered
continuously at 5 mL/h in ambient temperature of
25°C with new fully charged battery)
• Charging time:
≥
8 hours (when charged with AC power supply with
the power turned off)
The battery can be charged to approx. 80% in
3 hours.
Sub-battery: Nickel-metal-hydride battery (for a Power Failure alarm in case of
both power supplies are lost)
• Charging time:
≥
8 hours (when charged with a new battery and AC power
supply in ambient temperature of 25°C when the power is off)
* A state in which the equipment is unpacked and stored for subsequent use with the main power supply unplugged.