64
Changing Settings on the Menu
Increases value
Dial
Decreases
value
8)
Turn the Dial to set the Bolus VTBI time or the Bolus flow
rate.
Notes
• The Bolus VTBI time or the Bolus flow rate (mL/h) has been calculated from the
Bolus VTBI, and is displayed.
• The ready for bolus icon is displayed.
Ready for bolus icon
Bolus VTBI time
Bolus flow rate
• The Bolus flow rate can be set within the range of the following values.
Setting range:
0.01 to 150.00 mL/h (when using syringe of 5 mL)
0.01 to 300.00 mL/h (when using syringe of 10 mL, 20 mL, 30 mL)
0.01 to 1200.00 mL/h (when using syringe of 50/60 mL)
Step:
0.01 mL/h step (0.01 to 10.00 mL/h)
0.10 mL/h step (10.00 to 100.00 mL/h)
1.00 mL/h step (100.00 to 1200.00 mL/h)
• The Bolus VTBI time can be set within the range below.
Bolus VTBI time: 00min01s to 60min00s
Caution
• When using the Hands Free Bolus function, the setting cannot take a flow rate (mL/h) calculation result outside the flow
rate setting range, and therefore it should be reconfigured. [----.--] or [0.00] is displayed, not allowing solution delivery.
Purge switch
Operation indicator (green flashing)
9)
Press the Purge switch to start the bolus.
Notes
• The operation indicator flashes green.
• The bolus volume delivered increases.
• The alarm melody sounds during solution delivery.
Bolus volume delivered