Specifications
Power consumption
24VA
Classification
Electric shock
protection
Class I equipment and internally powered equipment, Defibrillation-proof type
CF applied part
Mode of operation
Continuous operation
Protection against
harmful ingress of
water or particulate
matter
IP24
Wireless LAN
Hereby, TERUMO CORPORATION declares that this TERUFUSION Syringe
Pump Type SS3 is in compliance with the essential requirements and other
relevant provisions of Directive 2014/53/EU.
Catalogue number: TE
*
SS830xxx*
• Transmit frequency or frequency band: 2412 to 2472 MHz
• Type of modulation and frequency characteristics:
• DBPSK (Differential Binary Phase Shift Keying)
• DQPSK (Differential Quadrature Phase Shift Keying)
• DSSS (Direct Sequence Spread Spectrum)
• CCK (Complementary code keying)
• Effective radiated power:
• IEEE802.11b 10.7 dBm
Catalogue number: TE
*
SS835xxx*
• Transmit frequency or frequency band: 2412 to 2472 MHz
• Type of modulation and frequency characteristics:
• DSSS (Direct Sequence Spread Spectrum)
• OFDM (Orthogonal Frequency-Division Multiplexing)
• Effective radiated power:
• IEEE802.11b 15dBm
• IEEE802.11g 13dBm
• IEEE802.11n 12dBm
* Alphanumeric characters are entered for x in the catalogue number.
Dimensions
Without external communication (RS-232C):
381 mm (W)*
1
x 120 mm (H)*
1
x 112 mm (D)*
2
With external communication (RS-232C):
381 mm (W)*
1
x 120 mm (H)*
1
x 131 mm (D)*
2
*
1
: Excluding protrusions
*
2
: Excluding protrusions, pole clamp and moving range
Unit Weight
Approx. 2.0 kg
Standard accessories
AC power cable, pole clamp (Screw type), instruction manual
• This product is in compliance with EMC (electromagnetic compatibility) standard IEC 60601-1-2:2014 (EN 60601-1-2:2015)
(CISPR group classification and class classification are Group 1 and Class A). It is also in compliance with the EMC level
required by IEC 60601-2-24:2012 (EN 60601-2-24:2015).
• Conformity standard and directive
IEC 60601-1:2005+A1:2012 (EN 60601-1:2006+A1:2013)
IEC 60601-1-2:2014 (EN 60601-1-2:2015)
IEC 60601-1-6:2010+A1:2013 (EN 60601-1-6:2010+A1:2015)
IEC 60601-1-8:2006+A1:2012 (EN 60601-1-8:2007+A1:2013)
IEC 60601-2-24:2012 (EN 60601-2-24:2015)
MDD (Medical Device Directive) 93/42/EEC (Class IIb)
Note
• If an AC power cable is not included with the pump unit, please contact TERUMO trained service technicians to receive the cable
suited for your region.