95
About External Communication Functions
Item
Size (Byte)
Details
Syringe status
1
0x0
None
0x1
(Reserved)
0x2
5 mL
0x3
10 mL
0x4
20 mL
0x5
30 mL
0x6
50 mL
0x7-0xFF
(Reserved)
Flow rate
3
Value of 100 times the set flow rate
Volume delivered
3
Value of 100 times the volume delivered
VTBI
3
Value of 10 times the VTBI
Infinite: 0xFFFFFF
Invalid value (when the function is disabled): 0xEEEEEE
Dosage rate
3
Value of 100 times the dosage
Unit of dosage rate
1
0x00: mL/h
0x04:
μ
g/kg/min
0x08: mg/kg/h
Weight
2
Value of 10 times the weight
Invalid value (when the function is disabled): 0xFFFF
Dilution
3
Value of 100 times the dilution
Invalid value (when the function is disabled): 0x000000
Unit of dilution
1
0x13: mg/mL
Invalid value (when the function is disabled): 0x00
Drug library Ver.
2
Version of the drug library
Drug library is registered: 0x0001-0xFFFF
Drug library is not registered: 0x0000
Drug library number
2
Number of the drug library
Selecting dose mode (weight mode) No.1 (
μ
g/kg/min): 0x0FFD
Selecting dose mode (weight mode) No.2 (mg/kg/h): 0x0FFE
Invalid value (when the function is disabled): 0x0000
Occlusion pressure
setting
1
bit0-bit3: Setting of occlusion pressure. 1 to 10.
Occlusion pressure range: 0x01-0x0A
bit7: Setting of occlusion detection pressure 1 = In 10 levels
(bit4-bit6: Reserved)
Hands On Bolus flow
rate
2
100.00 to 1200.00 mL/h (100 mL/h step)
Hands Free Bolus VTBI 3
Value of 100 times the Bolus VTBI
Invalid value (when the function is disabled): 0x000000
Hands Free Bolus dose
time (Bolus VTBI time)
2
0 to 3600 s (1 s step)
Invalid value (when the function is disabled): 0x0000
Hands Free Bolus flow
rate
3
Value of 100 times the flow rate
Invalid value (when the function is disabled): 0x000000
Syringe brand number
1
Brand number of the syringe selected
Total number of bytes
46
* For more details, contact TERUMO trained service technicians.