43
Other Operation Procedures
Confirmation
switch
7)
Check the parameter value and that “OK?” is selected,
press the Confirmation switch.
Notes
• Parameter changes can be identified by the change in background colour on the
parameter confirmation screen.
• When the Back/Mute switch is pressed on the parameter confirmation screen, the
parameter input is cancelled and the screen returns to that of step 3.
• Parameter values can be changed from the parameter confirmation screen.
• When the Confirmation switch is pressed, the flow rate display appears.
• Press the Display select switch to select info tab
. The setting information
appears.
• Parameter values can be changed from the screen displayed by selecting info tab
.
• For details about the contents that can be checked from info tab
, see
page 48.
Increases value
Dial
Decreases
value
8)
Turn the Dial to adjust the dosage.
For example:
μ
g/kg/min
For example: mg/kg/h
Flow rate
Dose rate Dose unit
Notes
• The dosage can be set within the range below.
Dose unit =
μ
g/kg/min: 0.01 to 999.00
Dose unit = mg/kg/h: 0.01 to 999.00
• Based on amount, solution, weight, dosage, the flow rate (mL/h) is automatically
calculated, and the flow rate display shows the result.
Caution
• If the calculation of flow rate (mL/h) is outside of the setting range*,
“0.00 mL/h” or “-----.-- mL/h” is displayed and solution delivery does not start
even if the Start switch is pressed. Make the setting again within the flow rate
setting range.
* For details about the flow rate setting range, see page 116.