94
About External Communication Functions
Item
Size (Byte)
Details
Alarm information
3
b0 (low order)
Shutdown Notice alarm
b1
Link Interruption alarm
b2
Completion alarm
b3
Occlusion alarm
b4
Syringe Displacement alarm
b5
Plunger Displacement alarm
b6
Slider Displacement alarm
b7
Syringe Barrel Detection alarm
b8
Nearly Empty alarm
b9-b12
(Reserved)
b13
Power Failure alarm
b14
(Reserved)
b15
AC power detector function
b16
No Flow Rate alarm
b17
Flow Rate/VTBI Volume Judgment alarm
b18
No VTBI alarm
b19
Battery alarm
b20
Start Reminder
b21
Pressure alarm
b22
Maintenance timer
b23 (high order)
Battery Replacement Time/Failure alarm
One bit used for each item. (“1” indicates active.)
Power status
1
b0-b3 (low order)
Battery level
0x0: (Reserved)
0x1:
Remaining battery level, 1 bar
0x2:
Remaining battery level, 2 bars
0x3:
Remaining battery level, 3 bars
0x4:
Remaining battery level, 4 bars
0x5:
Remaining battery level, 5 bars
0x6-0xF: (Reserved)
b4-b5
Power type
0x0: Battery
drive
0x1: AC
drive
0x2-0x3: (Reserved)
b6-b7 (high order)
Sub-battery status
0x0: (Reserved)
0x1: Sub-battery,
normal
0x2: Sub-battery,
abnormal
0x3: (Reserved)
Operation monitor
1
b0-b1 (low order)
Occlusion monitor (0x0: 0%, 0x1: 25%, 0x2: 50%,
0x3: 75%)
b2
(Reserved)
b3
Alarm sound status (0: Without sound, 1: With
sound)
b4
Keypad lock status (0: Not locked, 1: Locked)
b5
Plunger status monitor (0: Set, 1: Not set)
b6
(Reserved)
b7 (high order)
(Reserved)
One bit used for b2-b7. (“1” indicates active.)