Overview
Use infrared communication (IrDA), external communication (RS-232C) or wireless LAN to output the status of the pump to a
network system (e.g. monitor system or CIS). (The status of the pump cannot be changed from a network system.)
Ensure that the network system and the pump use the same communication conditions.
List of communication specifications
Communication
method
Infrared communication
(IrDA)
(TE-SS830 only)
External communication
(RS-232C)
(TE-SS732 only)
Wireless LAN
(TE-SS830 only)
Transmission rate
115200bps
115200bps
IEEE802.11b: 1Mbps to 11Mbps
IEEE802.11g*: 6Mbps to 54Mbps
IEEE802.11n*: 6.5Mbps to 72.2Mbps
Data length
8 bits
8 bits
-
Start bit
1 bit
1 bit
-
Stop bit
2 bits
1 bit
-
Parity check
None
None
-
Frequency (Europe)
-
-
2412 MHz (ch1) to 2472 MHz (ch13)
Encryption scheme
-
-
WPA2 (AES/TKIP), WPA (AES/TKIP), WEP
Operational mode
-
-
Infrastructure mode
LAN Standard
-
-
IEEE802.11b/g/n*
* IEEE802.11g/n is applicable to catalogue number: TE
*
SS835xxx only.
Alphanumeric characters are entered for x in the catalogue number.
Cautions
• When this product is connected with other devices, particularly note the following:
1. When connecting this product with other medical equipment or network system, check its conformance to IEC 60601-
1:2005+A1:2012 (EN 60601-1:2006+A1:2013) and IEC 60601-1-2:2014 (EN 60601-1-2:2015) prior to use in order to
ensure the system safety.
2. When connecting this product with a network system or other equipment, for safety, checking the manufacturer’s
specifications of the equipment is recommended. In addition, use EMI compliant products for the connection cable.
3. When using the external communication function, pay particular attention, as it is more susceptible to the effect from a
radio knife, mobile phone, radio device, defibrillator, etc. Regularly check that this product is operating normally.
4. If connecting with a nurse call, always check in advance with the electrical contractor for compatibility, and connect
with a system for which the connection function has been confirmed. [The functions regarding connection are not
guaranteed.]
5. If connecting with a buzzer or lamp instead of a nurse call, it should be used with a capacity of 12V DC, 1A or less.
6. When using the wireless LAN communication function, the effect on other equipment by radio interference should be
considered. [It may affect other equipment.]
7. When connecting this product with a network, the settings of this product and the network system need to be
compatible. Contact a vendor specialist for the correct settings and allow only the system administrator to perform
settings. [Failure to connect with correct settings may result in interfering with the original functionality or performance
of this product and affecting the network system.]
• Connection to a network including other devices may cause unpredicted and unacceptable risks for the patient, the user,
or a third person. Be sure to identify, analyze, evaluate, and control those risks.
• Subsequent changes to the IT-network could introduce new risks and require additional analysis.
- Changes to the IT-network include:
1. Changes in the IT-network configuration.
2. Connection of additional items to the IT-network.
3. Disconnecting items from the IT-network.
4. Update of equipment connected to the IT-network.
5. Upgrade of equipment connected to the IT-network.
About External Communication Functions