114
Troubleshooting
Flow rate
Occurrence
Cause
Action
• Pressing the Start switch does
not start solution delivery.
• The infusion line is not open.
• The syringe is not installed
correctly.
1. Replace the syringe correctly. (See page 26.)
2. Prime the line, ensuring that the plunger is kept in the
slider to prevent the solution from being delivered to
the patient. (See page 33.)
• An occlusion has occurred in
the infusion line.
• Eliminate the cause of occlusion. (See page 106.)
• Flow rate accuracy is low.
• Syringe has not been installed
correctly.
• Check the setting of the syringe. (See page 26.)
• The syringe in use does not
match the syringe brand set on
the pump.
• Check that the syringe brand setting matches the brand
of the syringe in use.
• The flow rate is not set
correctly.
• Correct the flow rate setting.
• There is large height
difference between the pump
and patient.
• Adjust the patient to the pump height differential.
External communication
Occurrence
Cause
Action
• Cannot communicate
(RS-232C)
• The baud rate, stop bit, parity,
or device ID number may not
be appropriate.
• Check the settings of the connected device.
(See page 87.)
• Cables are not securely
connected.
• Reconnect the cable. (See page 87.)
• Cannot communicate (IrDA)
• The infrared communication
window is dirty.
• Clean the infrared communication window with cotton
swabs or gauze, etc.
• The pump is not securely
attached to the TERUFUSION
Communication Rack System.
• Make sure that the TERUFUSION Communication Rack
System is securely attached to the pump.
See the instruction manual of the TERUFUSION
Communication Rack System.
• The TERUFUSION
Communication Rack System
is not turned on.
• Turn on the TERUFUSION Communication Rack
System.
See the instruction manual of the TERUFUSION
Communication Rack System.
• Cannot communicate (wireless
LAN)
• SSID, encryption setting, IP
address, subnet mask, default
GW, connection destination
IP address, communication
port, or device ID may not be
suitable.
• Check the settings of the product or the connected
device. (See page 87.)
• Signal strength icon (Colour =
Grey, No signal bar) flashing
• The pump is defective.
• Contact TERUMO trained service technicians.