15
Precautions
Warnings
• This product should not be used in conjunction with gravity infusion. [1. If this product suffers a downstream occlusion
lower than the connection with the gravity infusion line, the Occlusion alarm will not operate. 2. If bubbles are entrained in
the downstream of the infusion line due to emptied gravity infusion line first, the normal infusion cannot be carried out.]
• A syringe other than specified should not be used for this product. [If using a syringe other than those specified, the flow
rate accuracy and alarm functions cannot be guaranteed.]
Cautions
• When using a syringe (specified sterile syringe for general use) for this product, ensure that the nominal capacity of the
syringe is not exceeded. [If exceeded, the accuracy of flow rate or the alarm function cannot be guaranteed.]
• Use the AC power cable included and connect to an earthed AC power source. The AC power cable included should not
be used for other equipment. [Using an AC power cable other than the one specified may result in failure of this product. In
addition, if used without an earth connection, the electrical safety of this product is not guaranteed.]
• Before inserting an intravenous needle or connecting an infusion line with the connecting site, always bleed the air inside
the infusion line by pressing the purge switch and ensure that the slider pushes the syringe plunger. [1. This product
may cause harm to a patient due to air injection, since it does not have a function to detect air in the infusion line. 2. This
product may not be able to deliver solution for a while after started if there is a gap between the syringe plunger and the
slider, or between the syringe flange and the flange holder (on the clamp side).]
• Before starting solution delivery, check that the settings for any errors. (e.g. an error in the digit of flow rate etc.) [This
product does not have a function to determine the correct values, possibly resulting in excessive infusion or insufficient
infusion to a patient.]
• If the product is used under conditions that do not meet those specified for the flow rate accuracy described in
"Specifications", such as a low flow rate or short dosing time, the solution delivery should be monitored with extra care.
[The flow rate accuracy cannot be guaranteed.]
• Turn the power on before the syringe is inserted, and check that the LCD and operation indicator are flashing, and that the
buzzer sounds. [If the power is turned on when a syringe is installed, this product cannot operate a normal self-check (self-
diagnosis).]
• When installing a syringe, pull forward the clamp and turn. Then, after the syringe is set, turn the clamp back to the
original position and gently push it to fix the syringe. In addition, the slider should be moved while holding the clutch. [Any
unreasonable operation or excess operation will cause failure of this product.]
• When setting the plunger to the slider, ensure that there is no gap between the syringe plunger and slider. [If the plunger is
set and there is a gap, the air or drug solution may be sucked by the slider hook.]
• After a syringe has been installed, do not apply any force such as pulling or pushing the infusion line. [Applying these
forces may result in displacing the syringe barrel from the predefined position, temporarily delivering or sucking the drug
solution.]
• When starting solution delivery, check the volume delivered and then clear it as needed before use.
• When purging, check the purge volume as rapid infusion continues after the Purge switch is pressed. [It may result in
excessive infusion.]
• After an Occlusion alarm is issued, always eliminate the cause of occlusion before starting. [In order to correct the volume
of bolus at the time of handling occlusions, this product has functions to automatically reduce the pressure inside the
infusion line at the occurrence of the Occlusion alarm (pulling back the slider, subtraction from the volume delivered).
Therefore, if starting without removing the cause, the infusion may not be carried out correctly such as causing the drug
solution in the infusion line to flow back to the syringe, or a repetitive Occlusion alarm status.]
• This product should be placed and used on a level and stable location. In addition, when using the IV pole, ensure that the
stand is stable and this product is securely fixed to it. [Dropping or falling may result in damage or failure.]
• When fixing this product to the IV pole, the specified pole clamp should be used. [If a pole clamp that has not been
specified is used, the function cannot be guaranteed. It may cause failure or accident.]
• When attaching a pole clamp to the pump, ensure that it is set securely. [Failure to do so may result in falling of the pump
or pole clamp.]
• When a pole clamp is used, do not release it until you have checked that this product is securely fixed to the IV pole.
[Failure to do so may result in falling of this product due to the loosened pole clamp knob.]
• Disconnect the AC power cable from this product to remove it from the AC power source. Do not place this product in an
area with obstacles that block the AC power cable from being disconnected.