41
Other Operation Procedures
Selecting Dose Mode (Weight Mode)
The dose unit setting can be selected.
Note
• To disable the selecting dose mode, contact TERUMO trained service technicians.
Display
select switch
2 seconds
or more
1)
Press and hold the Display select switch for 2 seconds or
more while the solution delivery is stopped.
Note
• The menu 1 screen is displayed and then the dose mode select (Dose) display
appears.
Dose mode select tab
Confirmation
switch
2)
Press the Confirmation switch.
Note
• The existing dose mode (light blue) changes to the selected status (dark blue).
Display select switch
Dial
3)
Turn the Dial or press the Display select switch to select
“
μ
g/kg/min” or “mg/kg/h”.
Notes
• Turning the Dial or pressing the Display select switch moves the white frame.
• Select “mL/h” to return to the default setting.