Figure 6-6: Setup: Rocker window
2.
Touch
In Motion
.
Three data entry fields appear, and
In Motion
changes to black.
3.
Touch the Start Position field.
A data entry pad appears.
4.
Use the data entry pad to enter the desired number of degrees from the home position of 0°.
5.
Touch
Enter
on the data entry pad.
The Start Position field displays the position you entered.
6.
Touch the End Position field.
A data entry pad appears.
7.
Use the data entry pad to enter the desired number of degrees from the home position of 0°.
8.
Touch
Enter
on the data entry pad.
The End Position field displays the position you entered.
9.
Touch the Dwell Time field.
A data entry pad appears.
10.
Enter the desired dwell time using the data entry pad.
11.
Touch
Enter
on the data entry pad.
The Dwell Time field displays the number of seconds you entered.
12.
Touch
Confirm
.
The Setup: Rocker window closes.
Performing Tasks
6-12
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Содержание Quantum Cell Expansion System
Страница 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Страница 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Страница 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Страница 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Страница 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Страница 66: ...5 Operator Procedures Quantum Cell Expansion System Operator s Manual for Software Version 2 1 5 1...
Страница 73: ...Operator Procedures 5 8 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 74: ...6 Performing Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 6 1...
Страница 96: ...7 Set Management Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 7 1...
Страница 110: ...8 System Management Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 8 1...
Страница 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Страница 130: ...10 Load and Attach Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 10 1...
Страница 141: ...Load and Attach Tasks 10 12 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 142: ...11 Feed and Add Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 11 1...
Страница 155: ...Feed and Add Tasks 11 14 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 156: ...12 Release and Harvest Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 12 1...
Страница 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Страница 168: ...14 Quantum System Configuration Quantum Cell Expansion System Operator s Manual for Software Version 2 1 14 1...
Страница 171: ...Quantum System Configuration 14 4 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...
Страница 182: ...16 Maintaining the Quantum System Quantum Cell Expansion System Operator s Manual for Software Version 2 1 16 1...
Страница 185: ...Maintaining the Quantum System 16 4 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Страница 186: ...17 System Specifications Quantum Cell Expansion System Operator s Manual for Software Version 2 1 17 1...