Dedicated Controls
4–16
2715 User Manual
The Vert Scale controls select from the available logarithmic vertical scale
factors or convert to the linear mode.
100.0MHz
(AUTO SWEEP)
ATTN 6DB
23.8DBMV
VF WIDE
20.0MHz/
10 DB/
5MHz RBW (AUTO)
Channel Name CH nn
CALIBRATOR
10/5/1.
The VERT SCALE function block contains a three-way toggle key
labeled 10/5/1. This label expresses the three logarithmic vertical scale factors
available on the 2715: 10, 5 and 1 decibels per major vertical division. Press the
key to cycle the vertical scale factor through the three values in a 10 dB –5 dB
–1 dB –10 dB... sequence.
1. Press [10/5/1].
The signal peak is now one division down from the reference level. Also, the
vertical scale factor readout now indicates 5dB/. The noise also seems to
have disappeared; the signal-to-noise difference is the same, but the scale
factor change has moved the noise below the bottom of the screen.
2. Press [10/5/1] again.
The signal peak is now five divisions down and the readout says 1 dB/. The
primary use for this feature is to more accurately read displayed signal peaks.
3. Press [10/5/1] once again to restore the 10dB/ setting.
LIN.
The second key in the VERT SCALE function block is also toggle-action. It
converts the vertical scale from logarithmic to linear and back again.
4. Ensure that the calibration signal is centered and the reference level is set to
23.8 dBmV.
5. Press [LIN].
The vertical scale readout now indicates 1.94MV/. When the LIN mode is
initially selected, the 2715 converts the vertical scale so the bottom graticule
line is 0 V and the scale factor is converted from dB to volts. The display is
similar to what you see on an oscilloscope. Thereafter, the REF LEVEL
arrow keys change the scale factor in a 1-2-5 sequence. Consequently, the
reference level changes by 6 or 8 dB when changing scale factors. If the
FINE reference level step size is selected while in the LIN mode, the arrow
keys change the scale factor at a rate of approximately 0.02 division per step
and the corresponding reference level changes approximately 0.2 dB.
Vert Scale
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......