Spectrum Analyzer Menus
6–78
2715 User Manual
NOTE
. Any control settings can be used.
Many of the items on the Utility menu are intended for servicing of the 2715.
However, we will describe a few used to verify instrument performance. Items
not discussed are reserved for service personnel or for use under factory
supervision.
NOTE
. This procedure will overwrite saved waveforms without warning.
1. Press [UTIL] to call up the UTILITY menu.
2. Press [5] to select INSTR DIAGNOSTICS/ADJUSTMENTS.
3. Press [2] to select MANUAL ADJUSTMENTS.
4. Press [2] to select DISPLAY STORAGE CAL.
This selection creates a checkerboard display used to adjust trace position
and rotation. The trace rotation, vertical, and horizontal alignment controls
are located on the back of the 2715 (see Miscellaneous Controls on
page 4–33).
5. Turn the TRACE ROT control until the checkerboard test pattern is
rotationally aligned with the graticule.
6. Adjust the VERT POS control until the the top line of the pattern coincides
with the top graticule line.
7. Last, turn the HORIZ POS control until the vertical center line of the
display coincides with the center vertical line of the graticule.
8. After completing the alignment, repeatedly press [UTIL] until the spectral
display reappears.
You may improve the center frequency setting time in some applications by
changing the YIG oscillator settling delay time from the factory default value.
The factory YIG oscillator settling delay time default is 5000 ms. The indicated
YIG oscillator settling time (for example, the default of 5000 ms) represents the
longest delay that will occur. Typically, most operations of the spectrum analyzer
will result in shorter delay times. The default settling time was chosen to ensure
compliance with published specifications for all instruments over a variety of
conditions such as temperature, aging, wide frequency steps, calibration
variations, YIG oscillator variations, and other component variations.
All 2715 spectrum analyzer stability and frequency accuracy specifications are
guaranteed only at the 5000 ms setting. However, in many applications, faster
settling delay time may be successfully used. Determination of the suitability of
Diagnostics and
Adjustments
Aligning the Display
Changing the YIG Settle
Delay
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......