Spectrum Analyzer Menus
2715 User Manual
6–65
You can also store settings and waveforms by title as follows:
1. Using the Display menu, create a title for the current display (see Adding
Titles and Labels on page 6–36). Call it MY SETTINGS. Turn on title mode
and return to the spectral display.
2. Press [UTIL] [1] [4] [X].
This saves the settings under item 4 of the Stored Settings/Displays menu.
3. Press [UTIL] [1].
Item 4 on the menu is now MY SETTINGS. The settings are recalled like
any other, but the title often helps remind you what the settings are used for.
NOTE
. Any control settings can be used.
Normalization is a set of procedures contained in ROM that calculate gain and
frequency characteristics of the circuits in the 2715 based on the built-in
100 MHz, 18.8 dBmV calibrator signal. The characteristics must be accurately
known to correctly scale and display the analyzed signal. The normalization
values are stored in non-volatile memory (NVRAM). Using external signals to
achieve even greater measurement accuracy, or to renormalize the reference if
NVRAM should ever be lost is described later in Service Normalizations on
page 6–80.
If the 2715 self-test routine detects that the present 2715 gain or frequency
characteristics differ significantly from those determined during the previous
normalization, the NORMALIZATION SUGGESTED message is displayed on
the screen. You should then perform a normalization, although the 2715 remains
useable for relative measurements. The NORMALIZATION SUGGESTED
message may also appear while the 2715 is warming up, especially if the
user-defined power-up settings include a narrow span. This is normal, and you
need not perform a normalization unless the message persists beyond the
warm-up period. Any time you require the utmost accuracy, we recommend a
normalization first.
Normalization may be required because of any of the following:
H
Circuit or component variation with time (drift)
H
Large temperature difference from that at which previous normalization was
carried out
H
Nonvolatile memory (NVRAM) is lost for any reason
Normalizing the 2715
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......