Spectrum Analyzer Menus
2715 User Manual
6–7
This procedure enters an external attenuation value of 40 dB. The spectral
display does not change, but the reference level now indicates 58 dBmV and
is followed by the term OFST. This indicates that the reference level has
been offset, in this case by 40 dB. Any reference level now entered is offset
automatically by the external attenuation or gain.
28. Press [INPUT] [6].
Item 0 indicates that the offset is turned on.
29. Press [0] to turn off the offset.
The spectral display will reappear, but OFST is gone.
30. Press [INPUT] [6] [0] to toggle the offset back on without reentering the
external attenuation value.
31. Turn off the offset.
32. Enter a value of 0 dB for external attenuation.
The 2715 has a 75
W
input impedance and expects a 75
W
signal source
impedance. However, a 50
W
source impedance is typically associated with some
applications that use dBm as a standard amplitude measurement unit. The 2715
provides two ways to make 50
W
measurements:
H
When making narrowband measurements (for example, carrier-to-noise
ratios, relative amplitudes of signals close together in frequency, interference
levels relative to a nearby signal), you can generally connect the 2715
directly to a 50
W
source. If a 50
W
source is connected to the input of a
50
W
instrument, the voltage will be 1.6 dB lower than it is with the same
source connected to the 75
W
input of the 2715. We can also calculate that
0 dBmV dissipated in 75
W
is equal to –48.8 dBm across 50
W
. The total
difference is, therefore, –50.4 dB.
For these cases, use item 6 of the INPUT menu (EXTERNAL ATTEN/
AMPL) to enter a 1.6 dB offset to the reference level. See Accommodating
External Amplification/Attenuation on page 6–6.
H
When making broadband measurements or measurements of absolute
amplitude (for example, antenna, system, or amplifier sweeps; absolute
carrier amplitude; comparison of signals widely separated in frequency), you
may want to present a matched load to the source to provide maximum
flatness and to minimize standing wave ratios. You can do so by inserting the
matching minimum loss pad shown in Figure 6–3 between the source and
the 2715. To obtain correct dBm readings, set the INPUT menu measure-
ment parameters as follows:
Item 3:
REF LEVEL UNIT
DBM
Item 6:
EXTERNAL ATTEN/AMPL
–4
Accommodating a 50
Source
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......