Dedicated Controls
2715 User Manual
4–31
The center measure feature can perform still another function. When center
measure is used in the zero span mode, the counter measures the frequency at the
output of the 2715 detector. Because the output varies only if the signal is
modulated, the counter is actually measuring the frequency of the modulation (if
none is present, you receive a message saying so). Normally, AM detection is
used, but you can select FM detection. Whichever detector is selected, the center
measure feature provides a quick method of determining the frequency of the
modulating signal. Of course, the modulation frequency must fall within the
bandwidth of the resolution bandwidth filter and any video filter that is active
(and the FM discriminator when FM detection is selected).
Reset the 2715 controls according to the following settings box:
100.0MHz
(AUTO SWEEP)
ATTN 10DB
28.8DBM
VF WIDE
20.0MHz/
10 DB/
5MHz RBW (AUTO)
CALIBRATOR
The signal track feature of the 2715 continuously repeats the center measure
operation. On each sweep the signal nearest midscreen is remeasured and
recentered. This is useful for keeping a slowly varying or jittering signal centered
for close observation.
26. Press [CTR MEAS/TRKG] twice to activate signal track.
The term TRKG appears in center screen indicating that the 2715 is in the
signal track mode.
If you turn the tuning knob while a sweep is underway, the signal will be
displaced a bit during the next sweep. On the following sweep the signal is
recentered. The frequency readouts change because the tracking mode continu-
ously repeats the center measure feature.
If the signal being tracked falls below a preset threshold (see Setting the Signal
Threshold on page 6–22), tracking halts and the message NO SIGNAL FOUND
ABOVE THRESHOLD is displayed. The message is accompanied by a beep.
When the signal rises back above threshold, signal track automatically resumes.
27. Simulate signal loss and recovery by turning the calibrator off and then back
on again.
The audible beep makes this signal track mode useful as an amplitude
threshold detector.
You can also obtain continuous counter readings in the signal track mode.
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......