Spectrum Analyzer Menus
6–86
2715 User Manual
2. Change the resolution BW to 5 MHz.
3. Tune to a strong TV video carrier. Ensure the signal peak is close to the
reference level.
4. Call up the Sweep/Trigger menu and select SETUP TABLE, item 9.
5. Repeatedly select item 4, TV LINE STANDARD, until the TV line standard
appropriate to your signal appears. Use the OPEN (1024 line) setting for
nonstandard systems.
Items 0 through 3 of the SETUP TABLE determine which sync pulse(s) is used.
Choosing any of them also selects item 4, TV LINE mode, from the Sweep/Trig-
ger menu.
6. Select item 0, CONTINUOUS, and then return to the spectral display.
7. Turn off digital storage and select zero span.
8. Enter LIN mode, and change the sweep rate to 20
m
sec/div (sweep rate is
indicated in the left-hand readout column).
The screen should resemble Figure 6–16. In CONTINUOUS mode, the
sweep generator is triggered by the first pulse that occurs after the 2715
enters the ready-to-be-triggered state. In other words, after the 2715
completes the current sweep, it enters the ready-to-be-triggered state, and is
retriggered by the first horizontal sync pulse that comes along. Thus, a
different sync pulse generally triggers each sweep. The video signals
following each sync pulse are not quite the same, but are nevertheless
displayed on top of each other. The result is an intense, but fuzzy, display.
Figure 6–16: Video Signal Using Continuous Horizontal Line Triggering
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......