Spectrum Analyzer Menus
6–66
2715 User Manual
Always perform the normalizations in environments, especially temperatures,
like those in which the subsequent measurements will be carried out. Also,
remove any input signals from the 2715 before performing amplitude or
frequency normalizations. Although we recommend that you wait a full 15
minutes after power-up before making measurements with the 2715 (it is within
advertised specifications by then), it is possible to make them shortly after
power-up. If you must make measurements immediately after the instrument is
turned on and the NORMALIZATION SUGGESTED message is on screen,
carry out a normalization directly. Note, however, that if you do so, the 2715
may require renormalization after it is fully warmed up.
1. Press [UTIL] to call up the Utility menu.
2. Press [3] to select NORMALIZATIONS.
The Normalizations menu enables you to normalize amplitude parameters
(AMPLITUDE ONLY), the frequency parameters (FREQUENCY ONLY), or
both (ALL PARAMETERS). It takes longer for amplitude normalization than for
frequency normalization, and several minutes for both. Routinely select ALL
PARAMETERS to ensure the 2715 measures amplitude and frequency as
accurately as possible, but if you are in a hurry, you can select only the parame-
ters that are most important to your measurements.
3. Disconnect all signals from the 2715 input.
4. Press [0] to select ALL PARAMETERS.
Typically, you will see a number of changing waveforms and messages on screen
telling you which normalizations are being carried out. However, if NVRAM is
ever lost, the factory reference normalizations are lost with it (see If You Lose
NVRAM on page 6–81). Then, regardless of which normalizations you are
carrying out, you will receive the following message.
** ONE OR MORE REFERENCE NORMALIZATIONS HAS NOT
BEEN PERFORMED.
** FOR OPTIMUM ACCURACY, REFERENCE
NORMALIZATIONS MUST BE DONE
** CONSULT MANUAL FOR PROCEDURE
PRESS “W” TO CONTINUE
PRESS “Z” TO ABORT
5. If you have facilities to carry out the reference normalizations (see Service
Normalizations on page 6–80), abort this process and perform the reference
normalizations before continuing. Otherwise, press [W] to continue.
A recently normalized instrument is more accurate than one that is not normal-
ized, even if the references have not been normalized.
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......