Spectrum Analyzer Menus
6–80
2715 User Manual
The 2715 contains a set of frequency and amplitude normalization values when it
is shipped from the factory. The normalizations are based on reference values,
called reference normalizations, determined at the time of manufacture. The
reference normalizations specify the 2715 gain step sizes and the frequency and
amplitude of the internal calibrator signal. The references are determined by
comparison with an accurate external attenuator and signal source. If NVRAM is
ever lost (for instance, because the battery runs down), the factory reference
normalizations are lost with it, and you must perform a service reference
normalization to achieve maximum accuracy from the 2715. If you have signal
sources available that are more precise than the on-board calibration signal, they
can be used to achieve added accuracy when normalizing the 2715.
Using the REFERENCE NORMALIZATIONS selection from the Service
Normalizations menu, you can measure new reference values for the on-board
calibration signal and attenuator with respect to more precisely known alternate
sources. The alternate sources must exceed the following specifications:
frequency: 5 parts in 107,
±
10 Hz
amplitude: –18.8,
±
0.1 dBmV @ 100,
±
1 MHz
attenuation (gain step):
±
10 0.5 dB @ 100,
±
1 MHz
We will not give examples for all reference normalizations, but we will show you
how to determine a new frequency reference:
1. Connect an external frequency source to the 2715 input meeting the
frequency specification given above.
Various sources are available (for example, WWV in the U.S., some
broadcast television carriers, assorted frequency standards). A frequency of
100 to 500 MHz would be very good, but signals as low as 5 MHz can be
used.
2. Return to the spectral display and perform a center measure to ensure the
external reference is centered.
3. Press [UTIL] [5] [5] to select SERVICE NORMALIZATIONS.
4. Press [1] to select REFERENCE NORMALIZATIONS.
5. Press [1] to select INTERNAL REF FREQ from the resulting Reference
Normalizations menu.
On-screen prompts appear for each of the reference normalizations.
6. Press [1] to select EXTERNAL FREQ, and enter the frequency of the
external source.
Service Normalizations
Содержание 2715
Страница 4: ......
Страница 12: ...Table of Contents viii 2715 User Manual ...
Страница 18: ...Preface xiv 2715 User Manual ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Introduction 1 4 2715 User Manual ...
Страница 25: ...Getting Started ...
Страница 26: ......
Страница 46: ...Getting Started 2 20 2715 User Manual ...
Страница 47: ...Operating Basics ...
Страница 48: ......
Страница 70: ...Operating Basics 3 22 2715 User Manual ...
Страница 71: ...Dedicated Controls ...
Страница 72: ......
Страница 108: ...Dedicated Controls 4 36 2715 User Manual ...
Страница 109: ...CATV Measurements ...
Страница 110: ......
Страница 181: ...Spectrum Analyzer Menus ...
Страница 182: ......
Страница 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Страница 275: ...Appendices ...
Страница 276: ......
Страница 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Страница 318: ...Appendix C System Messages C 16 2715 User Manual ...
Страница 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Страница 339: ...Glossary and Index ...
Страница 340: ......
Страница 359: ......
Страница 360: ......